1 Principles of NIR spectroscopy
Molecules have several common quantized vibration and rotation states that can occur separately
or in combination.
A molecule with an electric dipole moment can go through one or more transitions
between various vibro–rotational states, if enough electromagnetic radiation at a specific
frequency is absorbed by the molecule. The energy levels of different vibration and
rotation states are quantized, and energy levels can be determined using the following
equation:
where:
- n is the vibrational quantum number
- h is the Planck’s constant
- f is a bond’s force constant, and
- μ is the reduced mass of the vibrating nucleus.
These transitions can occur with high probability for n equal to 1, whereas such absorption is
referred to as fundamental absorption when n = 1 or overtone when n = 2, 3,...,m, respectively.
The electromagnetic radiation absorbed for vibration state changes are in the infrared region, while
NIR spectroscopy utilizes absorption bands whose absorption is mostly due to overtones. NIR
instruments can usually operate with electromagnetic radiation wavelengths between 700 nm and
2500 nm, which are much shorter than that of NMR wavelengths (from 1 to 1000 nm).
Different molecules have varying constants of k and m, and thus, different NIR absorption
frequencies, or bands, in their spectrum. Oil has NIR absorption bands at: 890, 1162,
1720, 1760, 2308 and 2340 nm (for the ethylene groups). Proteins have NIR absorption
bands at: 1040, 1210, 1496, 2050 and 2140 to 2180 nm for the NH group. Water, or
moisture, has NIR absorption at the 960, 1150, 1405, and 1905 – 2000 nm regions for
the uncharged hydroxyl group. Sugars and carbohydrates have absorption bands at
2060 to 2150 nm, for their carbonyl group. This makes qualitative and quantitative
analysis feasible by NIR spectroscopy. The Lambert–Beer law states that the quantity of
electromagnetic energy absorbed by a sample is proportional to the amount of constituents in the
sample, and this allows quantitative analyses to be performed if the absorption can
be separated from other processes such as Light scattering and specular (mirror–like)
reflection.
Just as with visible light, NIR radiation can be reflected, transmitted and/or absorbed.
NIRS can be performed in either the reflectance or the transmission mode. As shown
previously, different constituents contain absorption bands at their specific corresponding
wavelengths.
1.1 NIR Techniques and instrumentation
NIR reflectance instruments have detectors that measure the intensity of the NIR radiation that is
reflected from the sample at several key wavelengths. The actual constituent contents can be
analyzed and calculated based on the calibration equation above, given the reflectance at the key
wavelengths.
However, NIR transmission instruments measure the intensity of NIR radiation transmitted
through a sample at several key wavelengths. A calibration equation is then created to relate “log
of reflectance” values at several key wavelengths to the actual constituent fractional content values,
usually done by comparing with wet chemistry analysis from a standard sample set using a
primary reference method. Since NIR transmission instruments measure the NIR portion of the
electromagnetic radiation that is actually transmitted through the sample, the path
length needs to be kept constant, and also selected for a high signal–to–noise ratio. NIR
spectroscopy instruments can also be referred to as discrete–region/filter systems, or
continuous spectrum detection systems, based on the mechanisms by which they separate
wavelengths. Discrete filter instruments select wavelengths by passing visible white light
(produced, for example, by a tungsten–halogen bulb) through a filter, allowing only a
predetermined, narrow region wavelength to pass through. Discrete filter instruments do
not collect data at all wavelengths, but only at or near the wavelengths of interest.
The biggest advantage of a discrete filter instrument is the high reproducibility of its
narrow wavelength ranges. The main limitation of a filter-based NIR instrument is that
absorption data is only collected at a few specified, narrow range wavelengths, and so the
initial wavelength range selection may be difficult if the sample matrix is unknown.
Filter-based, discrete wavelength instruments also tend to be slow if they are not utilized
in conjunction with simultaneous diode–array (DA) detection for several wavelength
ranges. Another limitation of these filter-based instruments is their limited spectral
resolution. For broad NIR absorption bands, the spectral resolution limitation may not be a
problem, especially if the selected filters satisfy the Nyquist spectral sampling criterion
[1].
Continuous NIRS instruments allow the collection of absorption figures for very large numbers of
wavelengths and thus, can be used to select or find wavelengths of interest for unknown matrices,
in order to develop a precise calibration method for a given type of analysis. Continuous spectra
systems can be divided into three subgroups: the moving grating/scanning systems, stationary
grating systems (which would include both diode array and AOTF systems) and Fourier transform
NIR spectrometers.
Moving grating, or scanning NIRS instruments utilize a moving grating to collect data at all
wavelengths, and as such, it is difficult to obtain reproducible scans, and the wavelength accuracy
also suffers. The stationary grating systems typically use parallel-processed diode arrays to collect
data from all wavelengths, and wavelength reproducibility and accuracy are thus significantly
improved. Another important advantage is that the scanning speed is also improved. A moving
grating system usually takes about half a minute to perform one scan, while the diode-array based
stationary grating system are claimed to be capable of collecting hundreds of spectra per second
[2].
Two modern NIR grain testing instruments, Zeltex models ZX800 and ZX50, were tested together
alongside the more sophisticated diode–array NIRS model Perten DA7000. The Zeltex ZX800 and
ZX50 are filter–based transmission NIR systems, with 13 and 14 filters, respectively. The
wavelength ranges for both of these instruments is from 893 to 1045 nm. The DA7000, on the other
hand, is a stationary grating instrument that allows the entire spectral range from visible light at
500 nm to NIR at 2500 nm to be acquired simultaneously in less than a second, using a
large array of silicon and InGaAs diodes. The ZX800 and ZX50 are both commercially
available from Zeltex (USA) with preliminary calibrations of protein, oil, and moisture for
soybeans, wheat and corn seeds. A total sugars calibration was also determined to be
possible with both these instruments [3]. The existing calibrations on the ZX800 and
ZX50 are primarily for ‘normal’, yellow coat seeds; calibrations for seeds with other
colors like brown or black are very difficult to create and usually inaccurate. Previous
research has revealed that NIR reflectance and transmission instruments have similar
accuracy and reproducibility [4, 5, 6]. However, the DA7000 is a relatively new type
of reflectance NIR instrument, and given that 86 pct. of the soybean hull consists of
carbohydrates, with only 9 pct. protein and 1 pct. oil [7], it could prove useful in analyzing these
non-standard grain samples. The analysis results from the ZX800 and DA7000 instruments were
compared to investigate the difference between transmission and reflectance instrument
accuracy as well as overall performance, especially when large sample sizes from different
seed varieties (or cultivars) are available for measurement. The DA7000 is distributed
commercially in the US by Perten Instrument North America Inc., but unfortunately
has no calibrations available for soybeans or corn seeds, with the exception of those
developed in our laboratory for these specific grains, as well as for green–colored soybeans
and 2 or 3 soybean seeds. The total sugars calibration for the ZX800 and ZX50, as
well as the calibrations for protein, oil, moisture and total sugars for the DA7000, were
based on our high–resolution NMR reference methods [8–19], as well as the primary
methods.
Three different models of FT–NIR spectrometers were extensively tested, and their performance
for food and grain applications was compared [1]. The first was a PerkinElmer SpectrumOne–NTS
instrument, and it was also the newest (2005/2006). The other two were the Bruker and the
Nicolet Technologies, NIR spectrometers. The SpectrumOne–NTS model, also commercially
available in the United States, appears to have both the hardware advantages of a well–designed
integrating sphere (NIRA), sample compartment, lower cost, as well as the ability to transfer
calibrations to similarly equipped models. Furthermore, the latter model is readily interfaced with
a room temperature, very high sensitivity Sb-detector for FT–NIR chemical/hyperspectral, or
microspectroscopic imaging, which is capable of approx. 1 μm resolution with about 10 pg
sensitivity. On the other hand, the other two NIR instruments tested proved to be more
versatile, flexible, and had much faster calibration software than the PerkinElmer model we
tested.
1.2 NIR Calibrations
NIRS instruments quantify protein content as well as the contents other components by measuring
the absorbance. The absorbance–log (1 = R) values are determined, and then related to the
fractional content of the component previously by a primary reference or standard method. The
process of establishing this relationship by utilizing a standard sample set is referred to
as a calibration. The association between the absorbance and chemical composition
is usually expressed as an approximate value and involves some forms of regression
equations.
According to the Lambert–Beer law, absorbance at a given wavelength is proportional to the
concentration of the component for solution of one component:
where:
- A is the (integral) absorption,
- k is the molecular absorption constant,
- l is the path length of the NIR light passing through the sample
- [c] is the concentration of the component responsible for the NIR absorption
(assuming that all scattering effects have been already corrected for!).
If a sample contains more than one absorbing component, then the absorption at a given
wavelength will be the total sum of the proportional contributions from all components in the
sample:
The measured absorbance is usually referred to as the apparent absorbance, and it
can be significantly affected by specular reflection and light scattering, even for thin
samples. Therefore, to obtain reliable NIR quantitation, spectral pre–processing and several
intensity corrections are always required. Thus, spectral variations between soybean
samples can be caused not only by chemical composition differences but also by spurious
effects that do not monitor chemical composition, such as specular reflection, multiple
scattering effects, and internal reflection. Our research group has found that the NIR
methods currently employed in industry for spectral pre–processing, including both
corrections for multiple light scattering and specular reflection effects, are in need of
substantial improvements to produce high accuracy, robust and stable calibrations for
rapid composition analyses of seeds. Our NIR calibrations were established mainly for
foods and whole kernel seeds, but calibration for ground samples were also successful.
Recently, protein content determinations were reported for single wheat kernels with
both transmission [5,21,22] and reflectance instruments [2,20]. Oil determinations for
single corn kernels were also reported with a transmission NIR instrument [23]. Wheat
single seed studies were also reported by NIR reflectance spectroscopy using the DA7000
instrument [2]. Calibrations for protein, oil, moisture, and total sugars of single soybean seeds
were obtained in our laboratory both with the DA7000 and the SpectrumOne-NTS
instruments.
The following are the principal NIR calibration steps:
- Generate or select a suitable set of standard samples of known composition
- Obtain raw FT-NIR data
- Correct NIR data for scattering
- Use Lambert–Beer law computations in conjunction with iterated data regression
by PLS–1 or PLS–2; also check up on specific PLS-1software packages for precision
and correct computation through numerical simulations for ideal testing, synthetic
numerical data
- Examine the calibration’s linear correlations, composition predictions, and validate
calibration with a wide range of unknown samples.
1.3 Regression methods
There are several regression methods that have been tested with NIR calibrations. Most
widely used regression methods include partial least square (PLS) [21, 24], principal
component regression (PCR) or principal component analysis (PCA) [5,24,25], and
multiple linear regression (MLR) [5,25]. The most widely used spectra pretreatment
method is multiplicative scatter correction (MSC) [24]. DA7000 has built–in software
able to carry our PLS, PCR, and PCA regressions and it also can enable or disable
MSC. The PLS, PCR, and PCA regression methods were all tried on DA7000, with or
without MSC correction. The best combination with the highest correlation coefficient is
adopted.